  6. 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V

3M™ Attest™ Super Rapid Readout Biological Indicator 1492V

3M ID 70200763566Catalog Number 1492V
  • Provides results in 1 hour
  • Offers 3X faster results than Attest Rapid BIs for 270°F and 275°F vacuum-assisted steam sterilization cycles
  • Helps deliver BI results in about the same time as it takes to cool a load
  • Meets ISO and FDA performance requirements for biological indicators
 More...
View all details
Specifications
Attribute Name
Value
Biological Indicator System
Super Rapid
Brand
Attest™
Cap Color
Brown
Catalog Number
1492V
Product Type
Biological Indicator
Product Use
Load Monitoring
Sterilisation Method
Steam
Product Details
  • Provides results in 1 hour
  • Offers 3X faster results than Attest Rapid BIs for 270°F and 275°F vacuum-assisted steam sterilization cycles
  • Helps deliver BI results in about the same time as it takes to cool a load
  • Meets ISO and FDA performance requirements for biological indicators
  • Contains viable organisms that provide a direct measure of lethality
  • Puts your instrument turnaround speed into overdrive for improved patient safety
  • *When used with 3M™ Attest™ Auto-reader 490 or the Attest™ Auto-reader 490H with software version 4.2.7 or greater.

1-Hour readout Biological Indicator for 132°C (270°F) 4-minute and 135°C (275°F) 3-minute Vacuum-Assisted Steam Sterilization cycles (Brown Cap)

The 3M™ Attest™ Super Rapid Readout Biological Indicator 1492V is a self‑contained biological indicator specifically designed for rapid and reliable qualification testing and routine monitoring of 270°F (132°C) and 275°F (135°C) dynamic‑air‑removal (pre‑vacuum) steam sterilization processes when used in conjunction with the 3M™ Attest™ Auto‑reader 490. The self‑contained design includes a carrier with spores of Geobacillus stearothermophilus and a media ampoule containing bacteriological media. The spore carrier and media ampoule are contained in a plastic vial topped with a brown cap. A chemical process indicator which changes from pink to light brown upon exposure to steam is located on the top of the cap. After processing, the biological indicator is activated and incubated in the 3M™ Attest™ Auto-reader 490. A final negative reading after 1 hour of incubation indicates an acceptable sterilization process.

Large Volume Order​

Pošaljite nam poruku

Hvala vam na interesovanju za 3M. Da bismo vam pomogli da efikasno upravljamo vašim upitom i odgovorimo na njega, ljubazno vas molimo da navedete neke ključne informacije, uključujući vaše kontakt podatke. Podaci koje navedete koristiće se za odgovor na vaš zahtev putem e-pošte ili telefona od strane predstavnika kompanije 3M ili jednog od naših ovlašćenih poslovnih partnera s kojima bismo mogli deliti vaše lične podatke u skladu sa politikom privatnosti kompanije 3M.

  •  
  •  
  •  
  •  
  •  
  • 3M takes your privacy seriously. 3M and its authorized third parties will use the information you provided in accordance with our Privacy Policy to send you communications which may include promotions, product information and service offers. Please be aware that this information may be stored on a server located in the U.S. If you do not consent to this use of your personal information, please do not use this system.

  • Slanje poruke

Hvala Vam što ste kontaktirali 3M

Primili smo vašu poruku i sada obrađujemo vaš upit
Neko od naših predstavnika će stupiti sa vama u kontakt putem telefona ili e-pošte

An error occurred.